Visual display system of changes in physiological states for patients with chronic disorders and data transmission via optical wireless communication
https://doi.org/10.17586/2226-1494-2022-22-5-951-961
Abstract
A visual display system of changes in the physiological state of patients and their data transmission via optical wireless communication is presented. Existing methods such as bedside monitors do not provide the possibility of visual display of information near the object in conditions of high workload of personnel and allow transmitting data via an electric cable or remotely in the radio frequency range. Such disadvantages reduce the responsiveness of personnel and lead to numerous errors in the transmitted data. A remote operational monitoring system is more effective, providing data collection from sensors, display with minimal delay time and data transmission over an interference-free communication channel in the optical range. A simple and visual way of displaying information allows to quickly and accurately recognize critical situations. The proposed system has two channels. The first channel provides a visual representation of changes in the physiological state on the visual display device in the form of a bracelet. The second channel allows transmitting data about the deviation in the patient’s condition via optical wireless communication in the infrared range of the spectrum for detailed display on a computer. The visual representation of changes in the physiological state is based on programmable change in the color of the LED indicators and on change in their operation modes. The block diagram and design of the visual display and data transmission system are presented. As part of the evaluation of the system operability, the heating of the visual display device in the COMSOL Multiphysics was evaluated. It is shown, that the average heating temperature of the indicator part in contact with human skin does not exceed 24 °C and is safe for the patient. The optical scheme of the receiving unit and the transmitting module is presented. The optical model of the module is presented in the Zemax program. It is shown, that the required average optical power of the working spectral region is 235 μW for the four infrared LEDs and four photodiodes located at a distance of 1 meter. A description of the working layout of a visual display and data transmission device is presented, including a digital pulse sensor and blood oxygen level, a color control and data processing unit, a visual display device, and a data transmission unit. The LEDs operation modes and its compliance with the deviation of the controlled parameters were checked by a special microcontroller program of practical significance. The implementation of the developed system is relevant when monitoring the condition of the object of observation in cases where the use of the radio frequency range is limited and the stability of the data transmission channel to electromagnetic interference plays an important role.
About the Authors
S. A. VostrikovaRussian Federation
Svetlana A. Vostrikova — Junior Researcher
Saint Petersburg, 197101
K. O. Pogorelova
Russian Federation
Kira O. Pogorelova — Engineer
Saint Petersburg, 197101
D. S. Shiryaev
Russian Federation
Daniil S. Shiryaev — Junior Researcher
Saint Petersburg, 197101
E. V. Tkacheva
Russian Federation
Ekaterina V. Tkacheva — Mentor
Saint Petersburg, 197101
N. A. Belyakov
Russian Federation
Nikita A. Belyakov — Junior Researcher
Saint Petersburg, 197101
Ya. N. Kovach
Russian Federation
Yakov N. Kovach — Engineer
Saint Petersburg, 197101
Yu. S. Andreev
Russian Federation
Yurii S. Andreev — Engineer
Saint Petersburg, 197101
I. G. Smirnova
Russian Federation
Irina G. Smirnova — PhD, Associate Professor, Associate Professor
Saint Petersburg, 197101
sc 57195684395
E. A. Kondratieva
Russian Federation
Ekaterina A. Kondratieva — D. Sc (Medicine), Head of MCS Study Group; Senior Lecturer; Senior Researcher
Saint Petersburg, 191014
Saint Petersburg, 194044
Moscow, 107031
sc 57191545581
References
1. Bakulin I.S., Kremneva E.I., Kuznetsov A.V. et al. Chronic Disorders of Consciousness. Ed. by M.A. Piradov. Moscow, Gorjachaja linijaTelekom Publ., 2020, pp. 1–12. (in Russian)
2. Balogh D., Kittinger E., Benzer A., Hackl J. Noise in the ICU. Intensive Care Medicine, 1993, vol. 19, no. 6, pp. 343–346. https://doi.org/10.1007/BF01694709
3. Busch-Vishniac I., West J., Barnhill C., Hunter T., Orellana D., Chivukula R. Noise levels in Johns Hopkins Hospital. Journal of the Acoustical Society of America, 2005, vol. 118, no. 6, pp. 3629–45. https://doi.org/10.1121/1.2118327
4. Сhakole M., Jibhkate R., Choudhari A., Gawali S., Tule P. A healthcare monitoring system using wifi module. International Research Journal of Engineering and Technology (IRJET), 2017, vol. 4, no. 3, pp. 1413–1417.
5. Eremenko A.A., Rostunova N.V., Budagyan S.A., Kurnosov A.V. The personalized telemedical system of Obereg for remote patient’s monitoring. Messenger of Anesthesiology and Resuscitation, 2020, vol. 17, no. 5, pp. 87–94. (in Russian). https://doi.org/10.21292/2078-5658-2020-17-5-87-94
6. Malika, Charu Rana. An indoor wireless Zigbee based patient monitoring system for hospitals. International Journal of Engineering Sciences Research-IJESR, 2013, vol. 4, no. 2, pp. 792–795.
7. Navya K., Murthy M.B.R. A Zigbee based patient health monitoring system. International Journal of Engineering Research and Applications, 2013, vol. 3, no. 5, pp. 483–486.
8. Mariappan P., Raghavan D., Abdel Aleem S., Zobaa A. Effects of electromagnetic interference on the functional usage of medical equipment by 2G/3G/4G cellular phones: A review. Journal of Advanced Research, 2016, vol. 7, no. 5, pp. 727–38. https://doi.org/10.1016/j.jare.2016.04.004
9. Dhatchayeny D., Sewaiwar A., Tiwari S., Chung Y. Experimental biomedical EEG signal transmission using VLC. IEEE Sensors Journal, 2015, vol. 15, no. 10, pp. 5386–5387. https://doi.org/10.1109/JSEN.2015.2453200
10. Ali H., Ibtehaj Ahmad M.I., Malik A. Li-Fi based health monitoring system for infants. Proc. of the 2nd International Conference on Communication, Computing and Digital systems (C-CODE), 2019, pp. 69–72. https://doi.org/10.1109/C-CODE.2019.8681012
11. Guger C., Allison B., Kammerhofer A., Guttmann F., Von Oertzen T.J., Spataro R., La Bella V., Annen J., Laureys S., Heilinger A., Ortner R., Cho W. MindBEAGLE — A new system for the assessment and communication with patients with disorders of consciousness and complete locked-in syndrom. Proc. of the IEEE International Conference on Systems, Man, and Cybernetics (SMC), pp. 3008–3013. https://doi.org/10.1109/SMC.2017.8123086
12. Wilheim F.H., Roth W.T., Sackner M.A. The LifeShirt: An advanced system for ambulatory measurement of respiratory and cardiac function. Behavior Modification, 2003, vol. 27, no. 5, pp. 671–691. https://doi.org/10.1177/0145445503256321
13. Zhang Y., Poon C.C.Y., Chan C., Tsang M.W.W., Wu K. A health-shirt using e-textile materials for the continuous and cuffless monitoring of arterial blood pressure. Proc. of the 3rd IEEE-EMBS International Summer School and Symposium on Medical Devices and Biosensors (ISSS-MDBS), 2006, pp. 86–89. https://doi.org/10.1109/ISSMDBS.2006.360104
14. Chan C.H., Zhang Y.T. Continuous and long-term arterial blood pressure monitoring by using h-Shirt. Proc. of the 5th International Conference on Information Technology and Application in Biomedicine, 2008, pp. 267–269. https://doi.org/10.1109/ITAB.2008.4570615
15. Lee B.-Ch., Kim J.Y., Choi S.H., Yoon Y.H. Use of an externalcooling device for the treatment of heat stroke. Clinical and Experimental Emergency Medicine, 2014, vol. 1, pp. 62–64. https://doi.org/10.15441/ceem.14.004
16. Gogate U., Bakal J. Healthcare monitoring system based on wireless sensor network for cardiac patients. Biomedical and Pharmacology Journal, 2018, vol. 11, no. 3, pp. 1681–1688. https://doi.org/10.13005/bpj/1537
17. Vostrikova S.A., Pogorelova K.O., Shiryaev D.S., Polukhin I.S., Andreev Y.S., Smirnova I.G., Kondratieva E.A., Bougrov V.E. Visual display system of changes in physiological state for patients with chronic disorders. Scientific and Technical Journal of Information Technologies, Mechanics and Optics, 2022, vol. 22, no. 2, pp. 401– 409. (in Russian). https://doi.org/10.17586/2226-1494-2022-22-2-401-409
18. Lovik R.D., Abraham J.P., Sparrow E.M. Surrogate human tissue temperatures resulting from misalignment of antenna and implant during recharging of a neuromodulation device. Neuromodulation, 2011, vol. 14, no. 6, pp. 501–511. https://doi.org/10.1111/j.1525-1403.2011.00396.x
Review
For citations:
Vostrikova S.A., Pogorelova K.O., Shiryaev D.S., Tkacheva E.V., Belyakov N.A., Kovach Ya.N., Andreev Yu.S., Smirnova I.G., Kondratieva E.A. Visual display system of changes in physiological states for patients with chronic disorders and data transmission via optical wireless communication. Scientific and Technical Journal of Information Technologies, Mechanics and Optics. 2022;22(5):951-961. https://doi.org/10.17586/2226-1494-2022-22-5-951-961